SET ID:
Reference:
Format:
Function:
number nn specified. If different from the current ID, the
instrument is immediately deselected. You must type Dnn
to bring it back on line.
The unit is selected using a selection code beginning with
the letter D followed immediately by a one or two digit
code in the range of 0-99. Further extraneous informa-
tion entered on the line is ignored. The first serial com-
mand sent to it should be Dnn to bring it on line. Ini-
tially, and when the instrument is re-initialized, it de-
faults to Device #1. The next return character sent to the
control will enable it as indicated by the MFCnn prompt,
where nn indicates the device ID. When the instrument
is powered up, it is deselected and must be enabled to
communicate.
Note: Although there are up to 99 different ID numbers
this does not mean that 99 can be driven off of one line. It
is suggested that you limit the number of units to no more
than 20 on a single RS232 line.
Example:
See Section 4-12 in Operator Manual.
SET ID (nn)
Where nn is a number from 1-99.
Sets the device ID of the instrument to the
SET ID 01
8-13
(CR)
(CR)