Adding Instruments to the GPIB Configuration

1. Select Add Instrument on the tool bar.
2. Select GPIB Interface > OK.
3. Select U3020A's Address in the GPIB address drop-down list > OK.
(Factory set address = 26)

Changing the GPIB Address

The U3020A GPIB address can only be changed programmatically. The GPIB address command
is: SYSTem:COMMunicate:GPIB:ADDRess <address>
When the address is changed, the new GPIB address is not updated in the Keysight Connection
Expert if the instrument was previously configured.
1. From the Keysight Connection Expert application window, highlight the instrument with
the address change and click Change Properties in the Configurable Properties
application window, change the address to the new address setting and select OK.

IP Address Command

This command assigns a static Internet Protocol (IP) address for the U3020A. Contact
your network administrator for the valid IP address to use for your instrument.
If you change the IP address, you must cycle power on the U3020A to
activate the new address.
Syntax
SYSTem:COMMunicate:LAN:IPADdress <address>
SYSTem:COMMunicate:LAN:IPADdress?
You can also queries the U3020A for the IP address it was assigned to.
Example
The following command sets the IP address:
SYST:COMM:IPAD 169.254.149.35
The following query returns the IP address currently being used by the instrument (quotes are
also returned).
SYST:COMM:LAN:IPAD?
Typical Response: "169.254.149.35"
Keysight U3020AS62 User's and Service Guide
Instrument Settings
17