6.2
System 2
–
Press the ON/OFF key
=>
Menu appears.
–
Press Ó,
System 2.
6.2.1 Updating module N 1700
!
The firmware file for the N 1700
module must first be copied to a
Micro SD card using a PC. The card
is then inserted into the C 1202.
–
Select the Flash module submenu
using the ▼ or
–
Select using
using the
Ó
–
Use ▼ or
▲
file for the N 1700 module in use, or
exit the submenu using the
!
Please check that the right firmware
file is saved on the Micro SD card.
If the wrong firmware is sent to
the N 1700 module, it will stop
working and will have to be sent to
Mahr Service.
–