3.6.2

Fitting the source enclosure to the instrument

Required material
Chemical-resistant, powder-free gloves
Warning:
To avoid personal contamination with biohazards or toxic materials, wear
chemical-resistant, powder-free gloves while performing this procedure.
To fit the source enclosure:
1.
Using both hands, slide the source enclosure onto the instrument's supporting rods.
2.
To secure the enclosure against the instrument, tighten the 2 thumbscrews on the front of
the source enclosure.
Notice:
them.
3.
Connect the electrical cable to the socket on the right-hand side of the instrument's front
panel, and tighten the screws.
Notice:
source enclosure.
4.
Fit the probe to the source enclosure (see
3.7

Maintaining the source components

Clean the source components when these conditions apply:
The sample cone and cone gas nozzle are visibly damaged.
You have dismissed inlet and sample-related causes for decreased signal intensity.
3.7.1

Removing the sample cone assembly from the ion block

Required material
Chemical-resistant, powder-free gloves.
Warning:
To avoid personal contamination with biohazards or toxic materials, always wear
chemical-resistant, powder-free gloves while performing this procedure.
To avoid damage to the electrical connector's screws, do not overtighten
To avoid damaging the fragile probe, use care when inserting it into the
June 28, 2016, 715003956 Rev. C
Page 30
page
22).