System settings to activate AutoCal
6
Waiting period until the measured value of the span gas is displayed on the screen
(HSS3).
7
Waiting period, RS1 AutoCal, until analyzer is in operating mode again.
4
5
6
7

Fig. 80: AutoCal activation: change system settings (second part of the list)

4
Option AutoCal test mode
6
Waiting period HSS3 span gas
After switching from manual calibration to AutoCal, the display changes to: 'Please re-start
the device OFF and ON again!'.
Note
The analyzer needs to be restarted after changing the parameters.
Start AutoCal procedure
Please tap on the calibration button. The M5/S1 screen opens. Scroll to 'AutoCal' on the
scroll bar at the left hand side.
PMA1000/PMA1000L | 1.00.00
5
Waiting period HSS2 zero gas
7
Waiting period RS1 AutoCal
www.mc-techgroup.com