These values are predefined. If the actual test gas concentration differs from these values,
please tap on the Edit button. The 'Channel K1 - Kn settings' with '3 = Calibration' screen
will open, where you can change parameters and values of the test gas concentrations.
To reach the AutoCal settings of this screen, please scroll down the
Note
parameter list.

Fig. 83: Parameter screen to adjust test gas measuring ranges and concentrations

After adjusting the parameters, please tap on the Calibration button to return to section
M5/S1. To start the AutoCal procedure, please tap on the green Start button.

Fig. 84: Tap on Start button to start AutoCal procedure

The automated calibration procedure starts. AutoCal begins with the zero gas calibration.
If you have also set a span gas calibration, this will follow as a second AutoCal procedure.
Info boxes display the current steps during the automated calibration process. The system
informs about waiting periods, and shows the actual switching states of the High-side and
relay outputs.
PMA1000/PMA1000L | 1.00.00
www.mc-techgroup.com
70