7.3.12 M4/S2 - Settings menu/ parameters

Qualified
Changing settings can only be done by qualified personnel.
personnel
In section M4/S2 you can define the parameters for the scroll bars you are using in section
M4/S1. The screen of section M4/S2 shows a scroll bar and a red 'Restart' button.
1

Fig. 38: M4/S2 screen with 'Restart' button

1
Scroll bar
After tapping on the 'Restart' button, a screen opens where you need to confirm the re-
start of the analyzer. The restart of the analyzer interrupts the measurement and deletes all
data collected during this day.
The RAM stores data collected from 12:00 a.m. until the next day at 12:00 a.m. After 24
hours of collecting data in the RAM, this data will be stored permanently in the flash mem-
ory of the analyzer. Any measuring values collected from 12:00 a.m. to the restart of the
analyzer will be deleted from the RAM.
Loss of data!
NOTICE
By tapping on the 'Restart' button, the measuring process is interrupt-
ed. The current measuring values in the RAM which are not perma-
nently saved, are lost.
With the scroll bar in section M4/S2 you can select different parameters. In the first range
there are 9 parameters and in the second range two, A and B.
To make sure that the settings will not be changed by accident, you will need to select the
parameter first by displaying it in the gray frame, and then tap on the 'hidden password'.
To select the parameters in the settings menu, please display the
Note
selected parameter in the gray frame of the scroll bar, and then tap
on the word 'Online' on the left-hand side of the system status line.
39
PMA1000/PMA1000L | 1.00.00
2
Restart-Button
2
www.mc-techgroup.com