Fig. 56: Display Of The High-Side- And Relay-Outputs With The Ma-Output Of I/O2; Fig. 57: Display Of The Highlighted Sm2 Components - M&C GENTWO PMA1000 Instruction Manual

Table of Contents

Advertisement

Fig. 56: Display of the High-Side- and relay-outputs with the mA-output of I/O2

Here all of the High-Side- and relay-outputs with the mA-output of I/O2 are displayed on
the left side of the screen. The switches are active, and you can test them by switching
them off ('0') or on ('1'). The mA-output can be changed by tapping on the displayed val-
ue. The keypad opens, and there you can enter the new output value. Please tap on the
'Execute' button to confirm your entry.
To check another part of the hardware, please tap on the module to get back to the M4/
S3 schematic screen.
You can also swipe horizontally to go back to the M4/S2 screen with the scroll bar. Display
'B=Diagnosis' in the gray frame of the scroll bar. Then tap on the hidden password again.
The screen on page 51 in Fig. 55 opens. Please tap on the hardware components to
select and highlight them.

Fig. 57: Display of the highlighted SM2 components

To test the internal data bus, please tap on the SM2 components. The screen displayed on
page 52 in Fig. 57 opens.
Please tap on the 'Test SM2' button to initiate the test. The line 'Connection check in prog-
ress' appears on the screen.
To return from the M4/S3 menu to the start screen, please swipe through the pages or tap
on the M&C button M1.
52
PMA1000/PMA1000L | 1.00.00
www.mc-techgroup.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Gentwo pma1000l

Table of Contents