Fig. 78: AutoCal activation: open section M4/S2

The screen with the system settings opens.
1
2
3

Fig. 79: AutoCal activation: change system settings (first part of the list)

1
Choice of zero, span or zero and span gas
3
AutoCal interval in hours
Enter settings
Please change the following system settings to activate the AutoCal functionality:
System settings to activate AutoCal
1
Choose your calibration gas: zero gas, span gas or zero + span gas.
2
Enter the time of the first AutoCal procedure. E.g. the current time is 13:25 and the
required start time for the AutoCal procedure 18:00. In this case please enter ' 4 '.
3
Enter the period the AutoCal procedure should be repeated. E.g. 24h means that the
procedure starts daily at 18:00.
4
Set the AutoCal test mode option to '0' (inactive)
5
Waiting period until the measured value of the zero gas is displayed on the screen
(HSS2).
67
PMA1000/PMA1000L | 1.00.00
2
Start time (24 hour format)
www.mc-techgroup.com