Gas Connections And Pin Assignment Diagram; Fig. 11: Gas Connections And Pin Assignment Diagram - M&C GENTWO PMA1000 Instruction Manual

Table of Contents

Advertisement

6.4 Gas connections and pin assignment diagram

Gas
Sample
Gas IN
P_IN
1
Sample
Gas OUT
2
P_OUT
Pressure
compensation
P_IN sensor pressure
∆_P used for flow signal
Power 100-240V, 50-60Hz
RJ45 Modbus TCP/IP
USB
Not used

Fig. 11: Gas connections and pin assignment diagram

The HSS-outputs for the AutoCal valves are only available for the
Note
PMA1000 analyzer.
The pin assignment diagram and the status signal RS1 use the following definitions:
Electronics:
HSS1...4
High-Side-outputs 1...4 (voltage level 0/24V, 500mA to connect
magnetic valves or coupling relays directly)
RS
Relays (potential-free relay output, max. 24V/2A)
P_IN
Pressure sensor input (measures the pressure compensation during
operation)
P_OUT
Pressure sensor input (measures the pressure compensation during
operation)
∆P
Pressure difference between P_IN and P_OUT
PMA1000/PMA1000L | 1.00.00
Electronics
SM 2
IO 2
SE2
X10
X10
X6
24V
SE1
X11
X2
X6
CAN
X9
X7
SE3
X7
Automated Calibration
(AutoCal, PMA1000 only)
IO 5
X10
X6
24V
X11
CAN
X9
X1
X2
X7
X3
X4
www.mc-techgroup.com
High side switches (HSS)
Measuring range
Short-circuit proof
feedback signal
24V DC / 0,5A
HSS1 = range 2
HSS2 = range 3
Operating parameters
1 2 3 4 5 6 7 8
X22
+
- + - + - +
-
HSS1 HSS2 HSS3 HSS4
HSS3 = Lim 1
HSS4 = Lim 2
mA
0/4...20mA
Analog Output
500Ω
X21
isolated
1 2
+
-
2x 24V / 2A
Status signal
X20
3 1 2 6 4 5
RS1= Alarm
RS1
RS2
RS2= Cal. Mode
High side switches (HSS)
AutoCal Valves
Short-circuit proof
24V DC / 0,5A
HSS1 = Sample/Test gas
HSS2 = Valve Test gas 1
X52
HSS3 = Valve Test gas 2
HSS4 = Valve Test gas 3
1 2 3 4 5 6 7 8
+ - + - + - +
-
HSS1 HSS2 HSS3 HSS4
X51
Not used
Pump Control
AutoCal Status
2x 24V / 2A
X50
RS1 = sample Pump On/Off
3 1 2 6 4 5
RS2 = Calibration Error
RS1