Manufacturer´s EMC declaration
The field strength of stationary transmitters such as base stations for portable telephones and mobile land radios, wireless
keyboard stations, AM and FM radios and televisions cannot be accurately theoretically predetermined. To determine the
electromagnetic environment with regards to the stationary transmitters, a study of the electromagnetic phenomena of the
location should be considered. If the measured field strength at the location in which the clTrac device will be used exceeds
the compliance level, the clTrac device should be observed to verify proper functionality. If unusual performance
characteristics are observed, additional measures may be required, e.g. modified layout or a different location for the clTrac
device.
Interference resistance
IEC 60601 test level
tests
Conducted disturbances
IEC 61000-4-6:2013
induced by RF fields
Radiated RF interference
IEC 61000-4-3:2006
variables in accordance
+A1:2007 +A2:2010
with IEC 61000-4-3
Compliance
level
3 V , 150 KHz to 80 MHz
6 V , ISM bands 150 KHz to 80 MHz
6 V , Amateur radio bands 150 KHz to 80 MHz
3 V/m , 80 MHz to 2.7 GHz, AM 1khz 80%
18