Zimmer CITrac Instructions For Use Manual page 25

Table of Contents

Advertisement

Operation Instructions
Patient interrupt button
Pausing the therapy
Notice:
Notice:
Notice:
During treatment, the patient can use the patient interrupt button. By clicking on the
patient interrupt button during a treatment a pop-up will show on the screen, accompanied
by an acoustic signal. The treatment will be paused, the force will be reduced to 10N.
By selecting the Stop button the treatment will stop and the Parameter screen will be
showed.
By selecting the Resume button the traction cord will build up it's tension and the program
will continue from the point of interruption.
Press the "Pause" button (15) to pause the therapy. The device will go back to the
parameter Settings screen. After making changes in the Parameter Settings select
"Resume". The treatment will continue from the point were the Pause button was pressed.
Only parameters in the current phase or upcoming phase(s) can be changed.
When pausing, the device will stop at its minimal set force.
When the treatment is stopped. All programmed information will be lost.