Error Messages
Troubleshooting
Disposal
Device malfunction
No response to the main switch / display remains dark.
Possible cause 1
Mains power connection
Remedy for cause 1:
Check whether the mains cable is properly plugged into the electrical outlet
and that the device connector is firmly inserted into the socket of the device.
Check the mains cable for damage. Replace it in the event of visible damage.
Check the power supply and mains socket.
Possible cause 2
Fuse
Remedy for cause 2:
In the power supply socket of the device, there are micro-fuses that disconnect
the power supply in the event of an electrical problem. Open the cover of the supply
socket and check the fuses. If necessary, replace the faulty fuse.
Only replace the fuse with a replacement that has the same name / is an exact
equivalent. Perform a complete check of the power supply for possible faults
beforehand.
If the fault persists, contact Zimmer MedizinSysteme GmbH or your local distributor
for help
Critical Error
Pop up Critical error occurred
Possible cause 1:
Electronics discovered a problem that cannot be solved by the user.
Remedy for cause 1:
Please contact Zimmer MedizinSysteme GmbH or your local distributor for help
No parameter selected
In the status / notification bar the text "No parameter selected" appairs
Possible cause 1:
The control knob is turned, while no parameter is selected.
Remedy for cause 1:
Select a parameter before turning the control knob