Error Messages
Troubleshooting
Disposal
Perform patient button
test to proceed
In the status / notification bar the text "Perform patient button to proceed" appairs
Possible cause 1:
The patient interrupt button is not yet pushed before starting treatment
Remedy for cause 1:
Press patient interrupt button to proceed the treatment
Possible cause 2:
The patient interrupt button is pressed but does not work
Remedy for cause 2:
Replace the patient interrupt button or contact Zimmer MedizinSysteme GmbH or
your local distributor for help.
Patient interrupt
button is not present
Pop up Patient interrupt button is not present
Possible cause 1:
Patient interrupt button is not connected to the device.
Remedy for cause 1:
Connect the interrupt button into the jack plug at the backside of the device
Possible cause 2:
Patient intterupt button is broken
Remedy for cause 2:
Replace the patient interrupt button or contact Zimmer MedizinSysteme GmbH or
your local distributor for help.