Error Messages
Troubleshooting
Disposal
End of rope detection
A pop-up appears notifying the user that they should release the rope and press the
button on the screen.
Possible cause 1:
The rope is pulled completely out of the device.
Remedy for cause 1:
Release the rope and press the button on the screen. The rope will be pulled
completely into the device.
Notice:
The confirm button is only available when there is no tension on the rope. Once the
patient is released, the confirm button is highlighted.
Notice:
If the rope is pulled during windup, the pop-up appears again. Please let go of the
rope to remedy and wait for the device to stop winding.