8 Troubleshooting
8
Troubleshooting
8.1
Possible Causes and Correction
Problem
Reduced positioning
accuracy
Function impairment
after system
modification
Mechanics do not
move; no operating
noise can be heard.
M-126 Precision Microtranslation Stage
Possible causes
When the M-126 is mounted
vertically: Load exceeds the
holding force of the drive
Increased wear due to small
motion over a long period of
time
Controller was replaced.
M-126 was replaced with
another model.
Controller and/or power
adapter are connected
incorrectly or defective.
Solution
Do not exceed the maximum permissible
loads according to the specifications (p.
33).
Do a maintenance run over the entire
travel range.
Controller from PI:
Load the parameters from the positioner
database that correspond to the
combination of controller and M-126
model.
Controller from a third-party supplier:
Check the operating parameters.
Check all connecting cables.
Check the controller.
If applicable: Check the positioner's
power adapter.
MP42E
Version: 3.0.0
29