7 Maintenance

7
Maintenance
7.1

General Notes on Maintenance

NOTICE
Damage due to improper maintenance!
Improper maintenance could lead to misalignment and failure of the M-126.
 Loosen screws only according to the instructions in this manual or the instructions of our
customer service department (p. 31).
7.2

Doing a Maintenance Run

Depending on the operating conditions and the period of use of the M-126, the following
maintenance measures are required:
Maintenance run
The maintenance run serves the purpose of distributing the existing lubricant.
 After 500 operating hours or at least after 1 year, do a maintenance run over the entire
travel range, in order for the existing lubricant to be distributed evenly.
 If you move the M-126 continuously over a short travel range (<20 % of the entire travel
range) in industrial operation, do a run across the entire travel range every 15 operating
hours or at least after 30 days.
Relubricating
Under laboratory conditions, it is only necessary to relubricate the M-126 in exceptional cases.
For continuous industrial use, the lubrication intervals must be determined individually.
 If you have any questions on relubricating, contact our customer service department (p.
31).