1.8

Name plate

Fig. 1: Example: Name plate for 265G, 265A transmitter
1 Device type/order code (for information
on the individual letters/numbers, refer to
the information on the order confirmation
or device data sheet)
2 Options – additional information on the
order code.
3 Device serial number (factory no.)
4 Identification code with reference to the
Pressure Equipment Directive (SEP or
1G). See the section "Compliance with
the Pressure Equipment Directive
(97/23/EC)".
5 Power supply
6 Parts that come into contact with medium
7 Lower to upper range limit (LRL to URL)
8 Set span
9 HART output (process variable)
10 Ex design designation (optional)
11 No. of EC-type-examination certificate
12 Tag (max. 32 characters)
IM/265Gx/Ax-EN-07
1
2
Type
265GS VEBNB1
Opt.
E1L1B2CBCLH3
3
S/N
265GS6502016232
4
PED: 1G
5
10,5 ... 30 V DC
Mat. 1.4404(8)/ AU/Hast.C
6
-1 ...+600 bar
7
eing./adj. +0.00 ... +50.000 bar
8
+0.00 ... +50.00 bar
9
II 1/2G Ex ia IIC T6
10
II 1D Ex iaD 20 T50°C or T95°C
Ta = -40°C to +40°C
11
ZELM 01 ATEX 0064
Electr. Data see EC-Type-Examination Certificate
TAG: -/-
12
265Gx, 265Ax
D-32425 Minden
/M
100%/+
0%/-
HART
SW xx
Bj./Yr. 20xx
PS: 600 bar
IP 67
4 ... 20 mA
/ L
min.Span 6.0 bar
or T4
or +85°C
13 EC conformity mark
14 Identification number of the notified body
with reference to the Pressure Equipment
Directive. See the section "Compliance
with the Pressure Equipment Directive
(97/23/EC)".
15 "SIL2" identification code (optional)
16 Device communication type
17 Software version
18 Year of manufacture
19 Ingress protection
20 Permissible pressure (static)
21 Output signal
22 Code for filling liquid
23 Minimum span
24 Identification number of the notified body
with reference to ATEX certification
(optional)
13
14
0045
SIL
2
15
16
17
18
19
20
21
22
23
24
0044
M00076
Safety
9