MEASURING FLUE GASES
After countdown is finished and your analyser is correctly set up, put your
flue probe into the appliance sampling point. The probe tip should be in the
flue centre - use flue probe depth stop cone to set position.
With balanced flues, make sure probe is positioned far enough into
the flue so no air can "back flush" into the probe.
SAFETY WARNING
Ensure your flue probe handle does not get hot!
Do not exceed analyser operating specifications - In particular:
Do not exceed flue probe maximum temperature (600°C)
•
Do not exceed analyser internal temperature operating range
•
Do not put analyser on a hot surface
•
Do not exceed analyser water trap levels
•
Do not let analyser particle filter become dirty and blocked
•
Check readings are stable and within expected range.
KANE958 MANUAL 17