SPECIFICATIONS
PARAMETER
Temperature Measurement
Flue Temperature
Inlet temperature
External sensor
Inlet Temperature
Internal sensor
(Internal Sensor)
Inlet temperature
Flue Gas Measurement
Oxygen
Carbon Monoxide H2
Compensated
Nitric Oxide (optional)
Nitrogen Dioxide
(optional)
Sulphur Dioxide
(optional)
Pressure Measurement
Pressure (Differential)
Calculations
CO/CO2 Ratio
Efficiency (Net or Gross)
Efficiency High (C)
Excess Air
Pre-programmed Fuels
UK, USA & France
European
Battery Life
Certification
RANGE
RESOLUTION
0 - 600°C
0 - 600°C
0.50°C
0 - 25%
0 - 10,000ppm
0 - 5000ppm
0 - 1000ppm
0 - 5000ppm
±160mbar
0 - 0.9999
0 - 99.9%
0 -119.9%
0 -119.9%
Natural Gas, LPG, Light Oil, Digester Gas, Coke, Wood,
Heavy Oil, Coal
>8 hours (continuous with pump on)
KANE958 is independently tested and certified to EN50379,
Parts 1-2 in accordance to 1st German Federal Emission
Control Ordinance (BimschV)
0.1°C
0.1°C
0.1°C
0.1%
1ppm
±5% > 400ppm - 2000ppm
±10% > 2000ppm -10,000ppm
1ppm
1ppm
1ppm
0.1mbar
0.0001
0.1%
0.1%
0.1%
KANE958 MANUAL 27
ACCURACY
±0.5°C
±0.5°C
±1°C
±0.3% Volume
±5ppm < 100ppm
±20ppm < 400ppm
±5ppm < 100ppm
5% > 100ppm
±5ppm < 100ppm
5% > 100ppm
±5ppm < 100ppm
±5% > 100ppm
±0.5% FSD
±5% of reading
±1% of reding
±1% of reading
±0.2% of reading