Full Roi Lut; Region Lut; User Defined Lut Settings; Predefined Lut Settings - Photon Focus MV1-D4096 Series User Manual

Table of Contents

Advertisement

7 Software
If LUT values should be retained in the camera after disconnecting the power,
then they must be saved with UserSetSave

7.5.2 Full ROI LUT

This section describe the settings for one LUT that is applied to the full ROI.
1.
Set LUT_EnRegionLUT (in category RegionLUT) to False. This is required to use the full ROI LUT.
2.
Set LUTEnable (in category LUTControl) to False. This is not mandatory but recommended.
3.
Select LUT 0 by setting LUTSelector (in category LUTControl) to 0.
4.
Set LUT content as described in Section 7.5.4.
5.
Turn on LUT by setting LUTEnable to True.

7.5.3 Region LUT

The Region LUT feature is described in Section 5.9.4. Procedure to set the Region LUT:
1.
Set LUT_EnRegionLUT (in category RegionLUT) to False. This is not mandatory but
recommended.
2.
Set LUTEnable (in category LUTControl) to False. This is not mandatory but recommended.
3.
Select LUT 0 by setting LUTSelector (in category LUTControl) to 0.
4.
Set properties LUT_X, LUT_W, LUT_Y and LUT_H (all in category RegionLUT) to desired
value.
5.
Set LUT content as described in Section 7.5.4.
6.
If two Region LUT are required, then select LUT 1 by setting LUTSelector (in category
LUTControl) to 1 and repeat steps 4 and 5.
7.
Turn on LUT by setting LUTEnable to True.
8.
Turn on Region LUT by setting LUT_EnRegionLUT (in category RegionLUT) to False.
7.5.4 User defined LUT settings
This section describes how to set user defined LUT values. It is assumed that the LUT was
selected as described in Section 7.5.2 or Section 7.5.3.
For every LUT value the following steps must be done:
1.
Set LUTIndex (in category LUTControl) to desired value. The LUTIndex corresponds to the
grey value of the 12 bit input signal of the LUT.
2.
Set LUTValue (in category LUTControl) to desired value. The LUTValue corresponds to the
grey value of the 8 bit output signal of the LUT.
The LUTIndex is auto incremented internally after setting a LUTValue. If consec-
utive LUTIndex are written, then it is required to set LUTIndex only for the first
value. For the next values it is sufficient to set only the LUTValue.
100 of 123
MAN069 12/2018 V1.1

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents