Helios PD2-6100 Strain Gauge, Load Cell, and mV Meter

Specifications

Except where noted all specifications apply to
operation at +25°C.

General

Display
Dual-line: 1.8" (46 mm) high, red LEDs
6 digits per line (-99999 to 999999), with
lead zero blanking
Display
Eight user selectable intensity levels.
Intensity
Default value is six.
Display
5/second (200 ms)
Update Rate
LED Status
See LED Status Indicators on page 23 for
Indicators
details.
Display flashes 999999
Overrange
Display flashes -99999
Underrange
Display Line 1:
Display
PV1, PV2, PCT, max & min, set points, PV
Assignment
& units, net & gross weight, Modbus input,
millivolts.
Display line 2:
Same as Display Line 1; plus units, tag or
turned off.
Units
Predefined: lb, kg, ounce, gram, ton, metric
ton (tonne); and custom units.
Programming
Four programming buttons, digital inputs,
Methods
PC and MeterView Pro software, or
Modbus registers.
Noise Filter
Programmable from 2 to 199
(0 will disable filter)
Filter Bypass Programmable from 0.1 to 99.9% of
calibrated span
Recalibration All ranges are calibrated at the factory.
Recalibration is recommended at least
every 12 months.
Max/Min
Max/min readings reached by the process
Display
are stored until reset by the user or until
power to the meter is cycled.
Select 1, 2, 5, 10, 20, 50, or 100
Rounding
(e.g. rounding = 10, value = 123.45, display
= 123.50).
Tare
There are three modes of tare operation:
Capture Tare, Programmable Tare, and
Off. See Tare Functionality (tare) on page
41 for details.
Three programmable passwords restrict
Password
modification of programmed settings.
Pass 1: Allows use of function keys and
digital inputs
Pass 2: Allows use of function keys, digital
inputs and editing set/reset points
Pass 3: Restricts all programming, function
keys, and digital inputs.
Non-Volatile
All programmed settings are stored in non-
Memory
volatile memory for a minimum of ten years
if power is lost.
Power
85-265 VAC 50/60 Hz; 90-265 VDC, 20 W max;
Options
12-24 VDC, 12-24 VAC, 15 W max.
Powered over USB for configuration only.
Fuse
Required external fuse: UL Recognized,
5 A max, slow blow; up to 6 meters may
share one 5 A fuse
Normal Mode
Greater than 60 dB at 50/60 Hz
Rejection
Isolation
4 kV input-to-power line
500 V input-to-output
(powered by external supply)
Overvoltage
Installation Overvoltage Category II:
Category
Local level with smaller transient
overvoltages than Installation Overvoltage
Category III.
Environmental Operating temperature range:
-40 to 65°C (-40 to 149°F)
Storage temperature range:
-40 to 85°C (-40 to 185°F)
Relative humidity:
0 to 90% non-condensing
Note: Printed circuit boards are
conformally coated.
Connections
Power, signal, relays, mA out: Removable
screw terminal blocks accept 12 to 22
AWG wire.
RS-485: Removable screw terminal block
accepts 16 to 30 AWG wire.
Digital I/O: Non-removable screw terminal
blocks accept 16 to 30 AWG wire.
Enclosure
UL Type 4X, IP65 rated. Polycarbonate &
glass blended plastic case, color: gray.
Includes four PG11 through-hole conduit
openings, with two factory installed PG11,
IP68, black nylon threaded hole plugs with
backing nuts.
Mounting
Wall Mounting:
Four (4) mounting holes provided for
mounting meter to wall. See Wall
Mounting Instructions on page 12 for
additional details.
Pipe Mounting:
Optional pipe mounting kit (PDA6260)
allows for pipe mounting. Sold separately.
See Pipe Mounting Instructions on page 13
for additional details.
Tightening
Power, signal, relays, mA out terminals:
Torque
5 lb-in (0.56 Nm)
Digital I/O and RS-485:
2.2 lb-in (0.25 Nm)
Overall
10.63" x 12.59" x 4.77"
Dimensions
(270 mm x 319.7 mm x 121.2 mm)
(W x H x D)
Weight
6.10 lbs (2.76 kg)
Warranty
3 years parts & labor. See Warranty
Information and Terms & Conditions on
www.predig.com
8
Instruction Manual
for complete details.