Setup And Programming; Overview; Led Status Indicators - Helios PD2-6100 Instruction Manual

Strain gauge, load cell, and mv meter
Table of Contents

Advertisement

Helios PD2-6100 Strain Gauge, Load Cell, and mV Meter

Setup and Programming

There is no need to recalibrate the meter when first received from the factory, if
used to read millivolts.
The meter is factory calibrated prior to shipment for millivolts with calibration
equipment that is certified to NIST standards.
If the meter is used with a strain gauge or load cell, it is recommended to perform a
live calibration with the sensor connected.

Overview

There is a 5-position DIP switch to set the meter input selection. See Switch Configuration on page 17 to setup the
switch.
Setup and programming is done using MeterView Pro or through the programming buttons.
After power and input signal connections have been completed and verified, apply power to the meter.

LED Status Indicators

LED
State
1-4
Steady
1-4
Flashing
1-4 & M
Flashing
T
Flashing
M
Flashing
G
Steady
Indication
Alarm condition based on set and reset points, independent of relay
status in certain configurations. (Available on all meter
configurations, including those without relays installed)
Relay interlock switch open
Relay in manual control mode
Meter in Tare mode
Analog output in manual control mode
Gross value being displayed
23
Instruction Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pd2-6100-6h0Pd2-6100-6h7Pd2-6100-7h0Pd2-6100-7h7

Table of Contents