Helios PD2-6100 Strain Gauge, Load Cell, and mV Meter
Dual-Scale Application
The mV input can be displayed in two different scales,
by enabling the dual-scale feature in the Setup Input
menu. See Dual-Scale (d-sCal) on page 28.
To enable the dual-scale feature you must select it in
the Input selection menu. See Setting the Input Signal
(Input) on page 28 for details.
Scaling the Input for PV1 (SCAL 1)
Scaling the Input for PV2 (SCAL 2)
Error Message (Error)
An error message indicates that the calibration or
scaling process was not successful.
After the error message is displayed, the meter
reverts to input 2 during calibration or scaling and to
input 1 during internal calibration, allowing the
appropriate input signal to be applied or programmed.
The error message might be caused by any of the
following conditions:
1.
Input signal is not connected to the proper
terminals or it is connected backwards.
2.
Wrong signal selection in Setup menu.
3.
Minimum input span requirements not
maintained.
4.
Input 1 signal inadvertently applied to calibrate
input 2.
Minimum Input Span
The minimum input span is the minimum difference
between input 1 and input 2 signals required to
complete the calibration or scaling of the meter.
Input Range
15 mV
25 mV, 30 mV
150 mV
250 mV, 300 mV
31
Instruction Manual
Input 1 & Input 2 Span
0.2 mV
0.4 mV
2.0 mV
4.0 mV