Troubleshooting Tips - Helios PD2-6100 Instruction Manual

Strain gauge, load cell, and mv meter
Table of Contents

Advertisement

Helios PD2-6100 Strain Gauge, Load Cell, and mV Meter

Troubleshooting Tips

This meter is a highly sophisticated instrument with an extensive list of features and capabilities. If the programming
buttons are used to program the meter, it may be a difficult task to keep everything straight. That is why we strongly
recommend the use of the free
the meter for programming with MeterView Pro software.
If you have programmed the meter with the programming buttons and it is not working as intended, try
re-programming the meter using MeterView Pro software.
Symptom
No display at all
Not able to change setup or
programming, Locd is displayed
Meter does not respond to input
change
Meter displays error message
during calibration (Error)
Meter displays
999999
1.
-99999
2.
fault
3.
Display is unstable
Display response is too slow
Display reading is not accurate
Display does not respond to input
changes, reading a fixed number
Display alternates between
Hi and a number
1.
Lo and a number
2.
Relay operation is reversed
Relay and status LED do not
respond to signal
Flashing relay status LEDs
Meter not communicating with
application programs
If the display locks up or the meter
does not respond at all
Other symptoms not described
above
Note: Certain sequences of events can cause unexpected results. To solve these issues, it is best to start fresh from
factory defaults and use the manual as a step by step programming guide, rather than a random approach to
programming. To reset the meter to factory defaults, see Reset Meter to Factory Defaults on page 50. In addition, for
best results, we recommend using the free MeterView Pro software for all programming needs.
MeterView Pro
software for all programming activities. A USB cable is provided with
Check/Action
Check power at power connector
Meter is password-protected, enter correct six-digit password to unlock or
Master Password of 508655.
If a Low-Value Cutoff has been programmed, the meter will display zero
below that point, regardless of the input – which can appear like the meter is
not responding to an input change. Check to make sure the problem is not
being caused by an undesired low-value cutoff setting.
To prevent the display from showing a negative value, set the low-flow cutoff
to a value greater than zero.
Check:
1.
Signal connections
2.
Input selected in Setup menu
3.
Minimum input span requirements
Check:
1.
Input selected in Setup menu
2.
Signal at Signal connector
3.
Input exceeds range selected
Check:
1.
Input signal stability and value
2.
Display scaling vs. input signal
3.
Filter and bypass values (increase)
4.
Increase Rounding value
Check filter and bypass values
Check:
1.
Input signal conditioner selected: Linear, square root, etc.
2.
Scaling or calibration
Check:
1.
Display assignment, it might be displaying max, min, or set point.
Press Menu to exit max/min display readings.
Check:
1.
Fail-safe in Setup menu
2.
Wiring of relay contacts
Check:
1.
Relay action in Setup menu
2.
Set and reset points
Relays in manual control mode or relay interlock switches opened.
Check:
1.
Serial adapter and cable
2.
Serial settings
3.
Meter address and baud rate
Cycle the power to reboot the microprocessor.
Call Technical Support for assistance.