Helios PD2-6100 Strain Gauge, Load Cell, and mV Meter
Sensor Excitation Voltage
Selection (EX+, EX-)
All meters, including models equipped with the
12-24 VDC power option, are shipped from the
factory configured to provide 10 VDC excitation
for the sensor.
If the sensor requires 5 VDC excitation, the
switch labeled P+/P- must be configured
accordingly.
To access the voltage selection switch:
1.
Turn off the power to the meter.
2.
Unscrew and open the lower door panel.
3.
Locate the P+/P- switch located in the
center of the connections board (see
diagram below).
4.
Flip this switch into the appropriate
position for the required transmitter
excitation.
MENU RIGHT
F1
MENU
ZERO

Figure 8. Sensor Excitation Voltage Selection

P+/P-
10V
5V
UP
ACK
F2
F3
RESET
TARE
15
Instruction Manual