Detailed Specifications

Normal-Sinus-Rhythm Waveform

ECG Reference ...................................................... The ECG amplitudes specified are for Lead II (calibration), from the
Normal Sinus Rhythm .......................................... 12-lead configuration with independent outputs referenced to right leg
High-Level Output ................................................. 0.5 V/mV ±5 % of the ECG amplitude setting available on a BNC
Amplitude .............................................................. 0.05 mV to 0.5 mV (0.05 mV steps); 0.5 mV to 5.0 mV (0.25 mV
Lead I .................................................................. 70
Lead II ................................................................. 100
Lead III ................................................................ 30
Lead V1 .............................................................. 24
Lead V2 .............................................................. 48
Lead V3 .............................................................. 100
Lead V4 .............................................................. 120
Lead V5 .............................................................. 112
Lead V6 .............................................................. 80
Amplitude Accuracy ............................................. ±(2 % of setting + 0.05 mV)
ECG Rate ............................................................... 10 BPM to 360 BPM in 1 BPM steps
Rate Accuracy ....................................................... ±1 % of setting
ECG Waveform Selection ..................................... Adult (80 ms) or pediatric (40 ms) QRS duration
ST-Segment Elevation .......................................... Adult mode only. -0.8 mV to +0.8 mV (0.1 mV steps)
Power-On Default .................................................. 60 BPM, 1.0 mV, adult QRS and ST-segment elevation of 0 mV

Pacemaker Waveform

Pacer-Pulse
Amplitude .......................................................... 0 mV (off), ±2 mV, ±4 mV, ±6 mV, ±8 mV, ±10 mV, ±12 mV, ±14 mV,
Accuracy
Reference lead II ............................................ ±(5 % setting + 0.2 mV)
All other leads ................................................. ±(10 % setting + 0.4 mV)
Pacer-Pulse Width ................................................ 0.1 ms, 0.2 ms, 0.5 ms, 1.0 ms, and 2.0 ms ±5 %
Paced Arrhythmias ............................................... Atrial 80 BPM
Power-On Default .................................................. Amplitude 10 mV, width 1.0 ms, atrial waveform
the FCC Rules.
baseline to the peak of the R wave. All other leads are proportional.
(RL). Output to 10 Universal ECG Jacks, color-coded to AHA and IEC
Standards.
connector.
steps). Other leads are proportional to Lead II (reference lead) in
percentage per:
Additional steps: +0.05 mV and -0.05 mV
±16 mV, ±18 mV, ±20 mV, ±50 mV, ±100 mV, ±200 mV, ±500 mV, and
±700 mV for lead II (reference lead)
Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
Atrio-Ventricular sequential
Noncapture (one time)
Nonfunction
Vital Signs Simulator
Detailed Specifications
85