Fluke Biomedical ProSim 8 User Manual page 71

Vital signs simulator
Hide thumbs Also See for ProSim 8:
Table of Contents

Advertisement

The pulse oximeter component of the device is not intended to
validate the SpO
2
The pulse oximeter component of this device is not intended to
confirm the SpO
2
oximeter monitor or to assess the optical characteristics of
representative pulse oximeter probes to determine their proper
calibration.
Not all functional testers and pulse oximeter equipment are
compatible. Functional testers can vary in pulse methods, pulse
contours, and amplitude. A functional tester might not
accurately reproduce the calibration of the pulse oximeter
equipment and can yield different results between pulse
oximeter equipment.
When the Type value is set to Masimo Rainbow, more SpO
parameters show on the display than other types of sensors. See
Test a Masimo Rainbow SpO2 to learn about these parameters.
To set oximeter SpO
2
show the screen in Figure 44 on the display.
Figure 44. Oximeter SpO
Connect the SpO
artificial finger to the SpO
2
Figure 45.
W Warning
accuracy of pulse oximeter equipment.
accuracy of the calibration curve of the pulse
Note
optical emitter and detector parameters, push  to
Optical Emitter and Detector Screen
2
Oximeter SpO2 Optical Emitter and Detector
2
jack on the front panel as shown in
2
Vital Signs Simulator
glh018.bmp
59