To set the ST Deviation:
1. Push or to move the highlight to the ST Deviation value.
2. Push .
3. Push or to adjust the deviation. Each push of a key moves the deviation
When the ST Deviation value is set to 0.00 mV, the value in the ECG
screen shows as Off.
4. Push to set the deviation and go back to the ECG screen.
To set the ECG signal amplitude:
1. Push or to move the highlight on the display to the Amplitude value.
2. Push .
3. Push or to adjust the amplitude. Each push of a key increases or
4. Push to set the signal amplitude and go back to the ECG screen.
Table 9 shows the percentage of the signal amplitude value that is put on each
ECG lead.
Waveform
Performance
Normal Sinus
Normal Sinus,
Horizontal Axis
Normal Sinus,
Vertical Axis
[1]
Includes R-Wave Detection, QRS Detection, and Tall T-Wave Rejection waves.
14
0.05 or 0.1 mV in the direction of the key pushed. The deviation range is
±0.80 mV
decreases the amplitude 0.05 mV between 0.05 mV and 0.50 mV and
0.25 mV above 0.50 mV to 5.00 mV in the direction of the key pushed.
Table 9. ECG Lead Amplitudes
I
[1]
70 %
100 %
70 %
100 %
100 %
30 %
30 %
130 %
Note
II
III
V1
30 %
100 %
100 %
30 %
24 %
48 %
-70 %
20 %
40 %
100 %
20 %
30 %
V2
V3
V4
100 %
100 %
100 %
100 %
120 %
112 %
70 %
100 %
80 %
40 %
60 %
80 %
V5
V6
100 %
80 %
50 %
100 %