You can go back a step when you push the Previous softkey. The Product
simulates that step for a full 15 seconds before it does the subsequent step.
When you push the Previous softkey while paused, the Product goes back a
step, but stays paused and sets the time to 15 seconds.
When you push the Stop softkey, the procedure simulation is stopped and the
initial test screen shown in Figure 19 shows on the display.
Simulate a Cardiac Catheterization Procedure
The Product simulates blood pressure measurements on both sides of a heart
valve. The pressure difference, or gradient across the valve, is used to determine
heart valve condition.
To simulate a Cardiac Catheterization procedure:
1. Push the Tests softkey in the IBP screen.
2. Push or to highlight Cardiac Catheterization.
3. Push . The valve selection shows on the display.
To change the valve value:
1. Push .
2. Push or to highlight Aortic, Pulmonary, or Mitral.
3. Push to go back to the IBP: Cardiac Catheterization screen.
To start the procedure simulation:
1. Push the Start softkey.
All the procedure steps are done manually.
2. Push the Insert softkey for the next step.
3. You simulate an increase and decrease of simulated pressure when you
push the Increase Pressure or Decrease Pressure softkeys.
4. Push the Pullback softkey to simulate the pressures when the catheter is
pulled back.
Push the Stop softkey to stop the procedure and go back to the IBP: Cardiac
Catheterization screen.
Invasive Blood Pressure Simulation and Tests
Vital Signs Simulator
35
Need help?
Do you have a question about the ProSim 8 and is the answer not in the manual?