Change RF Gen Amplitude, Output Port or Atten Hold (if on).
This message appears when the RF Generator's Amplitude field is set too high when
using the RF IN/OUT port or when adjusting the amplitude with the Atten Hold field
set to On.
The RF IN/OUT port has a lower maximum output level than the DUPLEX OUT port.
Use the DUPLEX OUT port, or reduce the RF Generator's level.
If Atten Hold is On, you may be adjusting the amplitude outside of the allowed range.
Change the amplitude
Decoder buffer full. Decrease gate time.
Too many decoder samples were sent to the decoder's buffer during a measurement gate
time causing a data overflow. Reducing the gate time decreases the amount of data sent
during each measurement.
Delta between RF Power and a channel level greater than 30 dB
A code channel level (pilot, sync, paging, or traffic) is set to a value outside of the 30 dB
range relative to the RF Power level.
To find out which code channel is causing this error message, calculate the delta between
RF Power and Sector A power. Add to this value the delta between Sector A power and
the code channel that is set to the lowest value. If this total exceeds 30 dB, decrease the
delta by adjusting the code channel level or the Sector A Power level closer to the
displayed RF Power. Repeat for Sector B if necessary.