Chapter 7, Description of Fields
Fields That Begin with the Letter S
Sensitivity
Serial No.
SID
262
This area performs two functions:
The upper field selects Normal or High sensitivity for the RF input. The High setting
adds about 6 dB of sensitivity to the ANT IN port for looking at very low level signals.
However, this setting can cause measurements to be uncalibrated (indicated by a mes-
sage on the screen). High sensitivity can also cause high-level AM signals to be dis-
torted.
The lower field selects the vertical resolution of the display. You can choose from
1 dB, 2 dB, or 10 dB per graticule.
Screen(s) Where Field is Present
SPECTRUM ANALZYER
CDMA REVERSE CHANNEL SPECTRUM
This field displays the serial number of the Test Set.
Screen(s) Where Field is Present
CONFIGURE
This field is used to set the system identification number (SID) of the Test Set.
The
field is an immediate action field. If the field is selected and changed, the
SID
signaling message(s) being sent on the forward control channel are immediately
changed. No change occurs to the current state (that is, active, register, page,
access, connect) of the Call Processing Subsystem.
This field is used to set the system identification number (SID) of the Test Set.
The :SIDentify command is used to control this field.
The query form of the command (that is, :SIDentify?) can be used to determine
the current system identification number (SID) setting.
Screens Where Field is Present
CALL CONTROL