HP 8924E User Manual page 176

Mobile station test set
Hide thumbs Also See for 8924E:
Table of Contents

Advertisement

Chapter 7, Description of Fields
Fields That Begin with the Letter D
De-Emphasis
Detector
176
This field is used to select or bypass the 750 uSec de-emphasis filter network used
to condition the audio signal before being analyzed by the Audio Frequency
Analyzer.
Screen(s) Where Field is Present
ANALOG MEAS
This field is used to select the type of detector used to measure the amplitude of
the audio signal being analyzed by the Audio Frequency Analyzer.
Detector Types
RMS displays the RMS value of signals.
RMS*SQRT2 displays the RMS value of a signal multiplied by
Pk+ displays the positive peak value.
Pk− displays the negative peak value.
Pk±/2 adds the positive and negative peak values, and divides the sum by 2.
Pk±Max compares the positive and negative peaks and displays the greater value
(polarity is not indicated).
Pk+ Hold displays and holds the positive peak value until the measurement is reset.
To reset, press the MEAS RESET key, select a different detector, or re-select the same
detector.
Pk− Hold displays and holds the negative peak value until the measurement is reset.
To reset, press the MEAS RESET key, select a different detector, or re-select the same
detector.
Pk±/2 Hold divides the sum of the positive and negative peak values by 2, and
displays the value until the measurement is reset. To reset, press the MEAS RESET
key, select a different detector, or re-select the same detector.
Pk±Mx Hold compares the positive and negative peaks and displays the greater value
until the measurement is reset. To reset, press the MEAS RESET key, select a different
detector, or re-select the same detector.
Screen(s) Where Field is Present
ANALOG MEAS
.
2

Advertisement

Table of Contents
loading

Table of Contents