RF Gen Freq
RF Gen Lvl
RF Gen Volts
This field sets the frequency of the CDMA generator. This field is displayed when
the
RF Display field on the CONFIGURE screen is set to Freq
See Also
RF Display field description, on page 250
Screen(s) Where Field is Present
ANALOG MEAS
This field is displayed only when the PCS Mode field is set to "On", an HP 83236
PCS Interface is installed, and power has been cycled to configure the Test Set for
PCS band testing. When this field is selected, the RF generator amplitude levels
are calibrated across the frequency bands made available by the HP 83236 PCS
Interface.
Operating Considerations
RF generator level calibration should be performed whenever operating
temperature changes more the 5 deg C, or when power is cycled to the HP 83236
PCS Interface, and before a call is set up. This calibration process takes
approximately 15 seconds.
Screen(s) Where Field is Present
CONFIGURE
This field specifies whether you want RF voltages expressed as the voltage across
a 50 ohm load, or the open circuit voltage (emf).
Operating Considerations
This setting affects the RF Generator's and the Tracking Generator's amplitudes.
Screen(s) Where Field is Present
CONFIGURE