Error −103
Error −104
Error −105
Error −108
Error −109
Error −110
Error −111
Error −112
326
Invalid separator
The parser was expecting a separator and encountered an illegal character. For example,
the colon used to separate the FREQ and AMPL commands should be omitted in the
following command:
RFG:FREQ 850 MHZ:;AMPL -35
Data type error
The parser recognized a data element different than one allowed. For example, numeric or
string data was expected but block data was encountered.
GET not allowed
A Group Execute Trigger was received within a program message (see IEEE 488.2, 7.7).
Parameter not allowed
More parameters were received than expected for the header. For example, the *ESE
common command only accepts one parameter; receiving *ESE 36,1 is not allowed.
Missing parameter
Fewer parameters were received than required for the header. For example, the *ESE
common command requires one parameter; receiving *ESE is not allowed.
Command header error
An error was detected in the header.
Header separator error
A character which is not a legal header separator was encountered while parsing the
header.
Program mnemonic too long
The header contains more than twelve characters (see IEEE 488.2,7.6.1.4).