Calibration
Digital Readout Message Codes
Listed below are Message Codes and a description of each which may be displayed on the Digital Readout during
the Calibration procedure.
DISPLAY
DESCRIPTION
MESSAGE
Indicates system is ready to process Digital Readout accuracy
C-1 OR C-2
calibration for a particular channel using the appropriate calibration
module.
CAL
Indicates that system is in the Calibration Mode.
Input to control circuitry unstable. Indicates that no calibration
E-5
module is connected to the channel being calibrated or the
incorrect module has been inserted.
Loose connection. Calibration input is out of range. Normally
E-6
occurs if incorrect calibration module is inserted.
H-X
Indicates the Current Channel is ready to accept new Upper
(X = 1 thru 9)
Temperature Limit setpoint X (X times 100° F).
L-X
Indicates the Current Channel is ready to accept new Lower
Temperature Limit setpoint X (X times 100° F).
(X = 1 thru 9)
OFF
This channel setpoint is below Lower Temperature Limit setpoint.
Indicates that theAutomatic Temperature Setback (and Power
S--
Down) feature is disabled (turned off).
Indicates that theAutomatic Temperature Setback (and Power
Down) feature is enabled (turned on) and will set each channels'
S-X
Set Tip Temperature back after X times 10 minutes of handpiece
inactivity (non-use).
---
Indicates that the system circuitry is proceeding with calibration
(flashing)
using the proper calibration module (C-1 or C-2).
REFER TO "CORRECTIVE MAINTENANCE" SECTION FOR OTHER ERROR
CODES
Table 2. Digital Readout Message Codes