Pace MBT250E-SDSoftGround Installation, Operation & Maintenance Manual page 42

Table of Contents

Advertisement

DIGITAL READOUT ACCURACY (CONT'D)
22. The system has now stepped to the next Active Channel. Repeat steps 12 through 21 to calibrate this
channel. If all channels have been calibrated, proceed to step 23.
23. Press and release the TIP OFFSET Key two times to exit Calibration Mode. All values, features and
defaults entered during the calibration are now stored in memory and all Set Tip Temperatures are
turned "OFF". All Channel Tip Temperature Offset values are set to the default value of "3" for °C ("6"
for °F).
42
Calibration