Pace MBT250E-SDSoftGround Installation, Operation & Maintenance Manual page 32

Table of Contents

Advertisement

PANEL CONTROLS (CONT'D)
18. The system will retain stored Set Tip Temperatures and Tip Offset Constants even when power is
removed.
19. Note the Current Channel displayed on the system. Turn the POWER Switch to the OFF ("0") position.
Turn the switch back to the ON ("1") position. Using the CH SELECT Key, select the channel displayed
at the beginning of this step.
20. Press the TIP OFFSET Key. Notice that the system has retained the stored Tip Offset Constants. Press
the key once again to exit Tip Temperature Offset Mode.
21. Press the TIP SET Key. Notice that the system has retained the stored Set Tip Temperature in memory.
Immediately press the TIP SET Key once again to exit Tip Temperature Set Mode.
22. In order to prevent a handpiece/tip combination from inadvertently operating at an incorrect Tip
Temperature, the system will not retain a stored Tip Offset Constant if a handpiece is disconnected.
The Tip Temperature Offset will return to the default value of "3" for °C ("6" for °F). Disconnect the
handpiece connected to the Current Channel. Reconnect the handpiece to the same channel. Notice
that the Current Channel changes to the next Active Channel.
23. Press the CH SELECT Key, as necessary, to change the Current Channel to the channel disconnected in
step 22.
24. Press the TIP OFFSET Key. Notice that the Tip Offset Constant has now changed to the default value
of "3" for °C ("6" for °F) and the Tip Offset LED turns off. Whenever a channel becomes inactive, the
system memory automatically reverts to the default.
25. While in Tip Temperature Offset Mode (Tip Offset LED illuminated), use the Scroll Up and Scroll Down
Keys to set Tip Offset Constant values as desired. Press and release the TIP OFFSET Key to exit Tip
Temperature Offset Mode and enter this value into system memory.
32
Operation

Advertisement

Table of Contents
loading

Table of Contents