°F/°C READOUT DEFAULT
6. Press and release the TIP SET Key. The Digital Readout will display "S - X" (X = "-" or 1-9). Either the
°F or °C LED will be on. This is the default temperature scale of the Digital Readout (e.g., if the °C LED
is on, the Digital Readout will display Tip Temperatures and Tip Offset Constant values in °C).
7. Press and release the °F/°C Key to change the default.
Each subsequent press and release of the key will change
the default.
AUTOMATIC SETBACK
8. As received from the factory, "S - -" will be displayed indicating that the Automatic Temperature
Setback is turned off. A "1" thru "9" appearing on the right side of the Digital Readout indicates time to
Automatic Setback in increments of 10 minutes. For example, "S-3" would indicate that any Active
Channel will set back its handpiece's Set Tip Temperature to 177°C (350°F) after 30 minutes of
handpiece inactivity (non-use). To change the time period or turn the Automatic Temperature Setback
feature off or on, use the Scroll Keys. Press the Scroll Up Key to increase the time period and/or enable
the feature. Press the Scroll Down Key to decrease the time period or disable the feature.
9. Press the TIP SET Key to store the °F/°C default and Automatic Temperature Setback time value in
system memory. The Digital Readout will revert to "CAL" and only the CH 1 LED will remain lit.
38
Calibration
Figure 20. Change Temperature Default