Mag-Prover-Self Diagnosis On/Off Setting - Toshiba LF516 Instruction Manual

Electromagnetic flowmeter capacitance type
Table of Contents

Advertisement

Switch operation
STEP3
STEP4(=END)
Note: If you try to set a value outside of the range −10% to +110%, LOW OVER SPEC or
8.2.24 Mag-Prover -Self Diagnosis ON/OFF Setting
You can select on/off setting for Mag-Prover's self-diagnosis function.
If the self-diagnosis function is set to OFF, no error message is displayed even if any of the errors listed
below occurs.
• ROM error
• RAM error
• System parameter error
• Excitation cable is not connected or its wiring is open
• Excitation circuit failure
• ADC circuit error
• Totalizer data is destroyed
OFF
ON
If this function is set to ON and an error occurs, an error message appears on the Display 2 measured
value screen. If an error occurs, the measurement items specified for Display 2 screen cannot be
displayed unless the error is removed.
Display example
H VAL
+103.0
%
H VAL
+103.0
SET OK?
ESC
O K
HIGH OVER SPEC error appears and the value returns to the previous value. Set a
value once again.
Selection items
Push
digit you want to change and push
to change the number of the digit.
When you push
to confirm whether the setting is OK or not.
If OK, push
setting, push
Pushing
ESC
and exits the setting screen.
N O
Description
Self-diagnosis function is turned off.
Self-diagnosis function is turned on.
− 92 −
6 F 8 A 0 9 3 4
Description
to move the cursor to the
, a message appears
. If you want to redo the
O K
.
N O
cancels the setting operation