13.2.4 When switch operation is unable

START
Does intense light shine on the
control panel such as direct sunlight
or reflected sunlight by window
glass or metal sheet?
Does the brightness change
always by ON/OFF of lighting?
Are you working with unfixed
lighting such as a flashlight?
Does the control panel get rained?
Is the control panel attached with
dewdrop, snow, ice, mud or oil?
Is the control panel hazed by
steam?
Are you wearing dark color gloves?
Or are your finger or gloves
stained?
Is there any light reflecting object
near the control panel?
Is there any reflecting object such
as metal plate placed opposing to
the control panel?
Contact your nearest Toshiba representative.
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
− 143 −
6 F 8 A 0 9 3 4
Place a cover to block the light from
surroundings, or cover the control
panel by hand to block the light.
Place a cover to block the light from
surroundings, or cover the control
panel by hand to block the light.
Operate the EMF converter fixing a
light to shine on the control panel as
constantly as possible.
Place a cover on the control panel to
protect from rain, and operate the
EMF converter.
Clean the attached or haze and operate
the EMF converter.
Further, place a cover to prevent from
attaching.
Clean the stain on fingers or gloves, or
wear gloves in light color and operate
the EMF converter.
Secure a space in front of the control
panel to allow a person to stand
Operate the EMF converter blocking
the light from the reflecting object by
standing in-between the control panel
and the reflecting object.
.