Digital I/O Specifications - Toshiba LF516 Instruction Manual

Electromagnetic flowmeter capacitance type
Table of Contents

Advertisement

10.1 Digital I/O Specifications

The specifications of the digital I/O terminals for the converter for electromagnetic flowmeter: LF546 are
as follows:
• Each I/O terminal can be used as a specified function terminal when selected.
• Terminal COM is the signal COMMON for the other three terminals (DO1, DO2 and DI).
• Each terminal is isolated from the internal circuits.
(The output terminals are not isolated from each other.)
Photo coupler
Digital Output 1(DO1 )
Output type:
Transistor open collector
Number of outputs:
1
Capacity:
30 V dc, 200 mA maximum
Digital Output 2(DO2 )
Output type:
Solidstate relay (non polarity)
Number of outputs:
1
Capacity:
150 V dc, 150 mA maximum
150 V ac(peal-to-peak), 100 mA maximum
Digital Input (DI )
Input signal:
20 to 30 V dc voltage signal
• High input level—20 to 30 V dc
• Low input level—2 V dc maximum
Input resistance:
Approximately 2.7 kΩ
Number of inputs:
One point
Solidstate relay
COM
DI
Voltage input (DI)
Contact output 2 (DO2)
Contact output 1 (DO1)
DO2
DO1
− 107 −
6 F 8 A 0 9 3 4
Transistor open collector
(Standard)
Resistor
Flowmeter