Toshiba LF516 Instruction Manual page 70

Electromagnetic flowmeter capacitance type
Table of Contents

Advertisement

Switch operation
STEP5(=END)
● Changing the span
You can set the span value for each range.
The following is an example to change the span of Range 1 from 2.0 m/s to 100 L/min.
Switch operation
STEP1
STEP2
STEP3
STEP4
STEP5
Display example
R TYPE
When you push
to confirm whether the setting is OK or not.
2F-2R
If OK, push
setting, push
SET OK?
Pushing
and exits the setting screen.
ESC
N O
O K
Display example
R1
Select "R1" from the setting item selection
menu.
2.00000
The currently set span value of Range1
m/s
(2.00000 m/s in this example) appears.
Then push
ESC
R1
The switches at the bottom change.
(
2.00000
At the same time, the cursor appears.
m/s
R1
Push
of the first unit.
2.00000
m/s
R1
Push
Push
3.93000
digit of the time unit.
L/s
R2
Push
Push
236.000
of span value.
L/min
− 69 −
6 F 8 A 0 9 3 4
Description
, a message appears
. If you want to redo the
O K
.
N O
cancels the setting operation
ESC
Description
.
are shown.)
to move the cursor to the digit
to change the first unit.
to move the cursor to the
to change the time unit.
to move the cursor to the digit