(2) Pipe Orientation
The detector may be installed in horizontal, vertical or sloping pipe runs as shown in
Figure 4.5. However, except for horizontal installation, fluid should flow from lower
to upper directions. See Figure 4.5.
Detector
(a)
The electrodes should be positioned horizontally against the ground surface in any piping
installation. See Figure 4.6.
Don't rotate.
Figure 4.6 Installation position of the detector
(b)
(c)
Figure 4.5 Detector Piping Orientation
Detector
− 22 −
6 F 8 A 0 9 3 4
Flow direction
(a) Horizontal pipe installation
(b) Vertical pipe installation
(c) Sloping pipe installation
Ground surface
Ground surface