3.1 C
TED 191 P
LEANING THE
The instrument itself should be cleaned using a damp cloth. In the event of an excessive
build-up of dirt, a mild alcohol solution (isopropyl) may be used to wipe over the instrument,
which should be allowed to air-dry afterwards. In extreme cases, the monitor casing may be
cleaned using a mild detergent.
Note:
The Accurox Oxygen monitor cannot be liquid sterilised or autoclaved –
irreparable damage may result.
3.2
S
TERILISING THE
The T-7 oxygen sensors and connecting cables used on the instrument may be gas
sterilised using low temperature ethylene oxide. A vacuum must not be drawn on the
sensor during the sterilisation process.
TELEDYNE ELECTRONIC TECHNOLOGIES
Analytical Instruments
O
ORTABLE
XYGEN
T-7 O
S
XYGEN
ENSOR AND
C
S
LEANING AND
TERILISATION
M
ONITOR
C
C
ONNECTING
ABLE
15