Teledyne TED 191 Instruction Manual page 19

Teledyne instruction manual portable oxygen monitor ted 191
Table of Contents

Advertisement

S
YMPTOM
The displayed readings
are unstable
Instrument reads "00"
Displayed value drifts
by 2 – 3% over 3 – 4
hours
Alarm continuously
activates when
displayed value is less
than 18%
TELEDYNE ELECTRONIC TECHNOLOGIES
Analytical Instruments
1) Remove the sensor tip from the T-adaptor, unscrew the "flow
diverter" and gently swab any excess moisture off the sensor cell
membrane, then flow dry gas over the membrane.
2) Check for a source of RFI emission. Relocate the instrument.
3) Check that the calibration control knob is properly locked in
position.
1) The sensor is not plugged in.
2) There is bad sensor/cable or cable/monitor connection.
1) Replace sensor cable.
2) Replace cable.
1) The minimum low alarm threshold is 18%. The Teledyne TED 191
will always alarm under these conditions.
T
ROUBLESHOOTING
C
/ R
URE
EMARK
19