Warning messages

Warning messages must be reset using the reset button in order to continue
working.
Figure 13: Reset button
Message
CLOCK ERR.
INI. APR
ERROR IN APR
TUBE SEASONING
HOT TUBE
EXPIRED TIME
MANUAL STOP
INACTIVE DAP
DR 100e (Digital Radiography) | Introduction | 33
Description
System clock error
APR checksum error
An APR value is out of
range
After a long idle period
(3 months or more) it is
necessary to perform X-
ray tube seasoning, to
avoid damage.
The temperature of the
X-ray tube has reached
the maximum allowed
value.
The exposure button
has been pressed half-
way for preparing the X-
ray tube for more than
15 seconds.
The exposure button
has been released be-
fore the end of the ex-
posure.
Dosimeter not connec-
ted.
Action
Press the reset button to
continue.
Press the reset button to
continue.
Change the value.
Press the reset button to
continue. Call Service
for the tube seasoning.
Wait for the monobloc
to cool down.
Release the exposure
button and repeat the X-
ray exposure sequence.
Press the reset button to
continue.
-