Equipotential Earth Connection - AGFA DR 100e User Manual

Digital radiography
Hide thumbs Also See for DR 100e:
Table of Contents

Advertisement

DR 100e (Digital Radiography) | Introduction | 45

Equipotential earth connection

The device is supplied with an equipotential earth connection point.
The device can only be used in areas that comply with local electrical safety
norms and in environments suitable for medical activities.
An additional equipotential earth connection is provided because the
equipment is movable and the reliability of the main equipotential earth
connection point can be insufficient.
It is only possible to use this equipment in rooms that comply with IEC norm
requirements.
0365B EN 20181015 0853

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6012/0016012/0026012/0036012/004

Table of Contents