Moving the unit

To bring the device to transport position and move it:
1. Make sure the device is switched off.
a) Press the power off button.
Figure 19: Power off
b) Unplug the power supply cable.
c) Wind the mains cable on the cable reel.
2. Move the rotating column in central position.
The two arrows (A) on the foot of the column are aligned.
DR 100e (Digital Radiography) | Basic Workflow | 65
CAUTION:
Risk of tilting during use or transport on inclined floors. Do not
use the unit on floors with an inclination higher than 5 degrees.
Do not drive the unit on ramps with an inclination higher than
10 degrees.
CAUTION:
The uncontrolled movement of the device could cause damages
to the operator, patient and personnel in proximity of it. The
equipment must be moved only after bring it to transport
position. For safe transport, switch off the equipment before
transporting it and ensure that all system peripherals (monitor,
mouse, keyboard, cables etc.) are disconnected.