102 | DR 100e (Digital Radiography) | Product Information

DR 100e

Electrical connection
Rated power supply
Standard outlet
Line resistance
Power supply cable
Infrared remote control
Radiological data
Nominal power (IEC 60601-1)
kV values
kV accuracy
mA values @115/230Vac
mA accuracy @115/230Vac
mAs values @115/230Vac
mAs accuracy
Exposure times @115/230Vac
Time accuracy
Max mAs
40 kV
41 - 45 kV
46 - 52 kV
53 - 62 kV
63 - 72 kV
0365B EN 20181015 0853
115 ÷ 230Vac ±10%, 50/60 Hz ± 5
Hz, 10A, standard monophase with
ground conductor.
Automatic setting of the equipment
according to mains voltage.
Automatic line compensation.
16 A @ 230Vac
<1 Ω @ 115/230Vac
8 m
2 x 1.5 V alkaline battery, 2500mAh
Battery life > 25000 exposures
32 kW @ 100 kV, 320 mA, 100 ms
40 - 125 kV at steps of 1 kV
±5% (IEC 60601-2-54)
50 - 400 mA
±10% (IEC 60601-2-54)
0.1 - 220 mAs
±10% (IEC 60601-2-54)
0.001 - 2.2 s according to mAs
±10% (IEC 60601-2-54)
max. 220 mAs
(the optional second capacitor bank is
required)
max. 200 mAs
max. 180 mAs
max. 160 mAs
max. 140 mAs